Published ahead of print on June 28, 2006
Clin J Am Soc Nephrol 1: 965-971, 2006
© 2006 American Society of Nephrology
doi: 10.2215/CJN.01090306
Relationship between Blood Flow in Central Venous Catheters and Hemodialysis Adequacy
Louise M. Moist*,
Brenda R. Hemmelgarn
, and
Charmaine E. Lok
* London Health Sciences CentreVictoria Hospital and University of Western Ontario, London, Ontario,
Foothills Medical Center and University of Calgary, Calgary, Alberta, and
University Health NetworkToronto General Hospital and University of Toronto, Toronto, Ontario, Canada
Address correspondence to: Dr. Louise M. Moist, London Health Science Centre-Victoria Hospital, 800 Commissioners Road E, Room A2338, London, Ontario, Canada, N6A 5W9. Phone: 519-685-8211; Fax: 519-685-8449; louise.moist{at}lhsc.on.ca
 |
Abstract
|
|---|
Central venous catheter dysfunction is a frequent problem and often is defined as a blood flow <300 ml/min. This prospective, cross-sectional study included 259 patients and examined the relationship between catheter blood flow and dialysis adequacy as measured by urea reduction ratio (URR), single pool urea kinetics, and online effective ionic dialysance clearance. Dialysis adequacy at blood flow rates of <300, <275, and <250 ml/min; sensitivity; specificity; and positive and negative predictive values were calculated. Mean blood flow was 352 ml/min (SD ± 48.8). Mean blood flow <300 ml/min occurred in 10.5% of the patients, and only 26% had a URR of <65%. Maximum blood flows <300 ml/min occurred in 6.9% of patients, and only 22.2% had URR <65%. The positive predictive value of mean blood flow of <300 and <275 ml/min to predict a URR <65% was 22 and 40%, respectively. Using receiver operator characteristic curves, the area under the curve was not significantly different for blood flows of 300, 275, or 250 ml/min. This study indicates that mean blood flows <300 ml/min are not commonly associated with dialysis inadequacy. Setting a single blood flow cut point of <300 ml/min to define the need for intervention will result in a significant number of unnecessary interventions. There is a need to reexamine the definition of catheter dysfunction and expand the definition beyond blood flow rates.
 |
Introduction
|
|---|
Cuffed, tunneled, dual-lumen central venous catheters (CVC) have become an acceptable form of hemodialysis (HD) vascular access when other permanent vascular accesses are unavailable, despite the associated high complication rates and mortality risk (14). In Canada, up to 70% of patients initiate HD with a CVC, and 33% continue to use catheters 90 days after dialysis initiation (5). This is far in excess of the National Kidney FoundationDialysis Outcomes Quality Initiative (NKF-DOQI) recommendation that CVC be used in <10% of prevalent HD patients (6).
A common complication of the CVC is its inability to deliver adequate dialysis as a result of dysfunction. The definition of CVC dysfunction varies across the literature and includes variable rates of blood flow ranging from <100 to 350 ml/min (716). The NKF-DOQI vascular access guideline defines CVC dysfunction as failure to attain and maintain extracorporeal blood flow that is sufficient to perform HD without significantly lengthening the HD treatment and considered insufficient extracorporeal blood flow to be <300 ml/min (6). This recommendation was opinion based and has been interpreted as the need to maintain blood flows >300 ml/min to ensure adequate dialysis. As a result of these guidelines, CVC often are run in the reverse configuration, thrombolytic agents such as thromboplastin inhibitor (t-PA) are used, or catheters are rewired when blood flow is reduced to <300 ml/min. With the increase in CVC prevalence and the corresponding increase in thrombolytic use and subsequent cost consequences, there is a clear need to clarify the relationship between CVC blood flow and dialysis adequacy. The primary objective of this study was to determine the relationship between mean blood flow in CVC and HD adequacy as measured by three parameters: urea reduction ratio (URR), single pool urea kinetics (spKt/V), and online effective ionic dialysance (EID) as measured by Diascan (Hospal Gambro, St. Leonard, Canada).
 |
Materials and Methods
|
|---|
Study Population
We undertook a prospective, cross-sectional study with participants who were recruited from two regional, university-based tertiary HD care programs at London Health Sciences Centre and the University Health NetworkToronto General Hospital. Research ethics board approval and individual patient consent were not required because this was undertaken as a quality improvement study and did not require additional investigations or alteration in delivery of care. Approval for access to patients was sought from the medical directors of each dialysis unit to collect the data as part of a cost management and quality assurance program, and the confidentiality of patients was maintained throughout the study.
Patients were included when they were on maintenance HD and using a cuffed-tunneled permanent CVC, located in the internal jugular vein, and were hemodynamically stable (defined as stable BP on dialysis without the need for saline bolus for the previous 2 wk). Patients received thrice-weekly, in-center dialysis using a single-use, high-flux polysulfone dialyzer. Patients at both centers underwent dialysis using the Integra (Gambro) dialysis machine. Patients at London Health Sciences Centre underwent dialysis using Permcath (Quinton Instrument Co., Seattle, WA) 12f caliber and 36 to 45 cm length, with a 2.5-cm separation between the arterial (proximal) lumen and venous (distal) tip. University Health Network patients used Uldall-Cook Catheter (Cook Canada Inc., Stouville, Canada), HIGHFLOW Dialysis Catheter (CardioMed Supplies Inc., Gormley, ON, Canada), Opti-flow/HemoGlide dual-lumen permanent dialysis catheter (Bard Access Systems, Mississauga, Canada), and Vaxcel Plus Chronic Dialysis Catheter (Boston Scientific, Boston, MA).
Assessment of Dialysis Adequacy
The Kt/V is a method of assessing the amount of dialysis that is delivered in terms of urea removal. This parameter was calculated using the Daugirdas formula Kt/V = ln (R 0.03) + [(4 3.5R) x (UF/W)], where UF is the ultrafiltration volume in liters, W is the postdialysis weight in kg, and R is the ratio of the postdialysis-to-predialysis urea (17). Patients with residual renal function were included in this study, but the renal clearance was not included in the Kt/V. The URR is a second method to estimate dialysis adequacy and was calculated by the equation (pre-urea post-urea)/pre-urea. Blood samples for the determination of plasma urea concentrations were drawn at the start and the end of the same dialysis as per the recommended guidelines using the slow flow method (18). Blood urea samples were analyzed using an enzymatic conductivity rate method (Synchron LX System; Beckman Instruments, Inc., Mississauga, Canada). The third measure of dialysis adequacy used the Diascan to measure the EID. EID uses conductivity measurements of both blood and dialysate to determine the clearance of sodium ions across the dialysis membrane. Several researchers have shown a relationship between EID and urea clearance and that it is possible to use ionic dialysance as an indicator of treatment efficacy (1923). The principle previously has been described extensively (24). The EID expresses clearance (K) as ml/min. K was used in the Diascan Kt/V calculation, where V was calculated as 60% of target body weight (kg) for men and 55% of target weight for women, and t was time on dialysis (in minutes). Dialysis treatment time varied according to the patient prescription. Dialysate flow rates were held constant at 500 ml/min during each treatment. Ultrafiltration rates depended on each patients requirements but were kept constant during sampling periods.
All blood measurements and Transonic monitoring was done during the midweek dialysis session concurrent with monthly blood work. Access recirculation was measured using the Transonic HDO1 monitor (Transonic Systems Inc., Ithaca, NY) (25). Three research nurses, trained in HD and in the use of the Transonic equipment, undertook data collection and Transonic measurements. The mean blood flow for that session (ml/min) was calculated by dividing the total liters processed by the number of minutes on dialysis. Demographics, cause of ESRD, comorbid conditions, and location of dialysis were collected at baseline. Adequate clearance was defined as a single pool (sp) Kt/V
1.2, URR
65%, and Diascan Kt/V
1.0.
Statistical Analyses
Mean and SD were calculated for continuous variables. Sensitivities, specificities, positive predictive values (PPV), and negative predictive values (NPV) of mean blood flow categories (<250, <275, and <300 ml/min) as an indicator of URR >65% were calculated. Receiver operator characteristic (ROC) curves were used to determine the relationship between blood flow rates and URR
65%.
Bivariate analyses were performed to evaluate the relationship between URR
65% and the following variables: Mean blood flow, catheter locking solution, percentage of recirculation, and line configuration. The bivariate analyses used a succession of independent sample t test for equality of means and repeated measures ANOVA (with categorical determinants), and correlation analyses (with continuous determinants). Variables whose coefficients were significant at P < 0.05 were included in a multivariate analysis to determine the relationship to URR. All statistical analyses were conducted using SPSS for Windows, Release 10.1.0, Standard Version (SPSS Inc., Chicago, IL).
 |
Results
|
|---|
A total of 259 patients completed the study. A subgroup of 89 patients had all three measures of dialysis adequacy (%URR, Kt/V, and Diascan Kt/V), whereas the remainder had measures of URR only. The baseline demographics and clinical characteristics of the 259 patients as well as the subgroup of 89 patients are presented in Table 1. The mean patient age was 66.0 ± 15.2 yr, and 51.0% were female. The percentage with hypertension, diabetes, and coronary artery disease were 79.5, 46.4, and 42.3%, respectively. The catheter and dialysis characteristics are described in Table 2 for both the total and the subgroup. The mean catheter age was 57.4 ± 61.6 wk with a range from 2 to 379 wk. The average duration of dialysis was 216.8 min, and the mean blood volume processed was 76.14 L. The calculated mean blood flow was 352.1 ± 48.8 ml/min. Recirculation was present in 108 (41.7%) of CVC; 28.7% occurred with CVC in the straight configuration, and 71.3% occurred in the reversed configuration. Recirculation of >10 and >20% occurred in 27.3 and 7.6% of the CVC, respectively. The catheter locking solution was heparin 1:500, heparin 1; 10,00, and 4% citrate in 32.3, 51.0, and 14.4% of lines, respectively. The remainders were locked with t-PA or were participants in the Pre-CLOT study (26).
The mean %URR for the total group was 73.8% (range 44 to 97%). The mean URR as well as the proportion of patients with a %URR <65 for each category of blood flow rate is described in Table 3. The mean URR for each blood flow category was not different (P = 0.07). Ten percent of patients had a URR <65%. Mean blood flow of <300 ml/min occurred in 27 (10.5%) patients, and 26% of these had URR <65%. Mean blood flow of >400 ml/min occurred in 11.3% of patients, and 13.8% of those had a mean URR <65%.
In the subgroup of 89 patients, the mean %URR was 74.6% (± 6.1), spKt/V was 1.65 (± 0.3), and Diascan Kt/V was 1.16 (± 0.23). The %URR, spKt/V, and Diascan Kt/V, categorized by blood flow rate, are presented in Table 4. The percentage of patients with URR <65%, spKt/V <1.2, Diascan Kt/V <1.0, and Diascan Kt/V <1.2 was 6.7, 5.9, 20, and 54.1%, respectively. Mean blood flow of <300 ml/min occurred in four patients, and only one patient had inadequate dialysis as measured by a URR <65%, spKt/V <1.2, and Diascan Kt/V <1.0.
Intervention to improve blood flow usually is based on a trend in blood flow or a single blood flow measurement versus a mean blood flow for the entire session. To explore this, we examined the maximal blood flow achieved during each dialysis session. The mean %URR and the proportion of patients with URR <65% for each category of maximal blood flow are listed in Table 5. The difference in mean %URR by maximum blood flow classification approached statistical significance (P = 0.055). Maximal blood flows <300 ml/min occurred in 6.9% of patients, and 18% had URR <65%.
We explored the diagnostic accuracy of mean blood flow to determine an adequate clearance as defined by URR
65%. We examined mean blood flow rates of <250, < 275, and <300 ml/min. The sensitivity, specificity, PPV, and area under the curve (AUC) in a ROC curve for mean blood flow rate are reported in Table 6. The PPV value, that is the proportion of patients who had a blood flow rate less than targeted when the URR indicated inadequate dialysis (URR <65%), increased from 22 to 40% when the target blood flow decreased from 300 to 250 ml/min. Lowering the target blood flow from 300 to 250 ml/min resulted in a small, NS decrease in NPV from 91.5 to 90%.
The AUC indicates the overall predictive accuracy of a test. A test with perfect predictive accuracy has a combined AUC of 1.0, whereas a test with no predictive accuracy has an AUC of 0.5. The AUC for blood flows of 250, 275, and 300 ml/min were 0.53, 0.55, and 0.57 (P = 0.89).
Patient age, gender, mean blood flow, percentage of recirculation, catheter age, catheter locking solution, and predialysis weight were examined in a bivariate analysis to examine the association with %URR. Line configuration and maximal blood flows were not included because of the close correlation with percentage of recirculation and mean blood flows, respectively. The following variables were significantly associated with increased %URR: Lower predialysis weight (kg; P = 0.001), higher mean blood flow (ml/min; P = 0.012), lower percentage of recirculation (P = 0.001), female gender (P = 0.04), and increased patient age (P = 0.04). Catheter capping solution was not associated with dialysis adequacy. Using variables that were significant at P < 0.05 in the bivariate analysis and including dialysis time, several variables continued to be associated with increased URR, including higher mean blood flow (P = 0.001), female gender (P = 0.001), longer dialysis time (P = 0.001), and increased age (P = 0.004). As expected, a higher dialysis weight (P = 0.001) and increased recirculation were associated with lower %URR. Altogether, these six independent variables explained a significant variation in %URR using the following regression equation with inclusion of standardized ß coefficients in order of decreasing influence on %URR: URR = 50.3 0.476 (predialysis weight) + 0.288 (mean blood flow) + 0.223 (if female) + 0.222 (time) 0.202 (recirculation) + 0.147 (age).
The clinical interpretation of this is as follows: Each increase of 1 kg in predialysis weight results in a decrease of 0.476% in URR. Each increase of 1 ml/min in mean blood flow results in an increase of 0.288% in URR. Female patients URR averages 0.223% higher than male patients URR. Each increase of 1 min in session length results in an increase of 0.222% in URR. Each increase of 1 yr in patient age results in an increase of 0.147% in URR.
 |
Discussion
|
|---|
Current vascular access guidelines recommend an intervention in a CVC when one fails to attain and maintain extracorporeal blood flow that is sufficient to perform HD without significantly lengthening the HD treatment. The guidelines then state a blood flow of <300 ml/min could be considered insufficient (6). The results of this study would challenge the validity of this recommendation. In this study, we found that 10.5% of dialysis sessions had a mean blood flow of <300 ml/min, and 75% of these achieved adequate dialysis as defined by a URR
65%. When we examined other measures of dialysis adequacy, only one of four patients had inadequate dialysis with mean blood flows <300 ml/min. Even when we looked at the proportion of patients with a maximal blood flow of <300 ml/min throughout the dialysis session, only 22% of patients had inadequate dialysis. A mean blood flow cut point between 250 and 300 ml/min had low predictive accuracy in determining a dialysis session with a URR <65%. Last, we have shown that several variables play a role in determining dialysis adequacy. Interventions to maintain adequacy, based on blood flow alone, will result in an inappropriate increase in intervention and cost.
CVC are used in an increasing proportion of patients who are on HD (2,5). The reported incidence of catheter dysfunction ranges from 5 to 80%, depending on the definition, population, and catheter characteristics (716). The interpretation of studies that use blood flow as a measure of catheter dysfunction are challenging without knowledge of blood pump setting, use of average versus episodic measures of blood flow, percentage of recirculation, and other dialysis variables. There is an increasing use of long daily, nocturnal dialysis and slow low-efficiency dialysis, all of which use lower blood flows (27,28). Defining dysfunctional catheters by blood flows of <300 ml/min in this population would lead to unnecessary intervention in the majority of patients.
Few, if any, studies have defined catheter dysfunction as the inability to deliver an adequate dialysis prescription, which is the overall goal of dialysis. The availability, acceptance, and further validation of online measurements of EID may aid in describing the dialysis dose at each session and allow for identification of inadequate dialysis in a more timely and precise manner. In this study, we empirically used a Diascan Kt/V >1.0 as the measure of adequate dialysis after we identified that >50% of the patients had a Diascan Kt/V <1.2. This Diascan Kt/V target has not been validated with clinical outcomes. We chose a lower Diascan Kt/V target because EID has been shown consistently to underestimate delivered dialysis dose, even after correcting for recirculation and urea distribution volume (1929). Further studies using EID clearance and its relationship to blood flow and adequacy using a CVC need to be performed.
We calculated the sensitivity, specificity, PPV, and NPV of mean blood flow as a diagnostic tool to predict dialysis adequacy in CVC. A mean blood flow of <300 ml/min had a PPV of 22%, whereas a mean blood flow <250 ml/min had a PPV of 40%. This indicates that at a mean blood flow of <300 ml/min, only 22% of patients had a URR <65%, whereas 40% had a URR <65% at a blood flow of <250 ml/min. The tradeoff between sensitivity and specificity was examined using the ROC curves. The AUC was not significantly different for the blood flow cut points of 250, 275, and 300 ml/min.
There are significant clinical and cost implications of defining catheter malfunction on the basis of blood flow alone. The misconception that blood flows <300 ml/min lead to inadequate dialysis often results in nursing personnels reversing the line to maintain flows >300 ml/min. Carson et al. (30) recently described the relationship among blood flow, line configuration, percentage of recirculation, and clearance. Despite documented recirculation, it is possible to achieve acceptable clearance if the blood flow is sufficiently high; however, there are tradeoffs, and some patients will have unacceptable clearances in the reversed configuration (30). Patients may be subjected to unnecessary line changes and increased intervention with t-PA. The costs that are attributed to the use of t-PA and line exchange are not known. A recent cost analysis of vascular access calculated CVC at $9180 Can (median $3812; interquartile range $2250 to $7762) per patient-year (16). In this study, catheter dysfunction was defined as blood flow of <200 ml/min, and the cost of t-PA was $199 per patient-year at risk. Increasing the intervention target to a blood flow of 300 ml/min would significantly increase costs. In our institutions, the cost of 2 mg of t-PA to each port is approximately $120.00 Can. If all patients in this study who had a blood flow <250 ml/min were treated (n = 5), then the cost of t-PA for that session would have been $325.00 Can. If we intervened in all lines with a mean blood flow <300 ml/min, as per K/DOQI recommendations, then we would have treated 28 lines at a cost of $1820.00 Can. The cost of t-PA would have increased five to six times from changing the target blood flow for intervention from 250 to 300 ml/min. Importantly, this use would be justified if mean blood flow <300 ml/min correlated with an important clinical consequence, such as inadequate dialysis; however, our data suggest that 22 (80%) of 27 lines would have been treated unnecessarily at an excess cost of $1365.00 for a single dialysis session.
Dialysis adequacy was associated with patient variables age, predialysis weight, and female gender, as well as dialysis variables time, mean blood flow, and percentage of recirculation. Although this is well-established knowledge in nephrology (31,32), it serves as a reminder that multiple variables influence the delivery of adequate dialysis dose, in addition to blood flow.
The question, then, is which blood flow rate should be used to assist in determining the need for intervention to maintain adequacy? Clearly this cannot be established for a general HD population. For a 40-yr-old, 100-kg man, a mean blood flow of 300 ml/min or more may be needed to maintain adequacy, whereas an 80-yr-old, 55-kg woman may have adequate clearance with a mean blood flow of 275 ml/min. The recent Canadian Hemodialysis Clinical Practice Guidelines on vascular access have defined CVC dysfunction as the inability to maintain pump speed >200 ml/min for one run or difficulty aspirating from either lumen of the catheter (33). This may be a more conservative estimate of the blood flow cut point. None of the patients in our study had blood flows <200 ml/min, so we were not able to examine the relationship between dialysis adequacy and this blood flow cut point. On the basis of the trend of our data, blood flows <200 ml/min would result in inadequate dialysis in the majority of patients.
Our study has a number of limitations. We used a calculated mean blood flow that was based on the liters processed and time on dialysis, which some consider inferior to recorded average blood flow from the blood pump or the gold standard of blood flow as measured using ultrasound dilution techniques (3436). Interventions in CVC usually are based on an intermittently low blood flow or inability to withdraw blood at the initiation of dialysis versus mean blood flow as used in this study. To address this, we examined dialysis sessions with maximum blood flows <300 ml/min and found that 82% of these sessions provided adequate clearance, thus confirming our original results. Given the cross-sectional nature of our study, we were unable to examine trends in blood flows and dialysis adequacy.
The strengths of our study include the use of a large number of dialysis patients from two separate sites, which increases the generalizability of the results. We were able to correlate various levels of blood flow with three separate and complementary measures of dialysis adequacy.
 |
Conclusion
|
|---|
In this study, mean blood flow rates <300 ml/min were not commonly associated with dialysis inadequacy. Setting a single blood flow cut point of <300 ml/min to define the need for intervention will result in a significant number of unnecessary interventions and associated increased costs. There is a need to reexamine the definition of catheter dysfunction and expand the definition beyond blood flow rates.
 |
Acknowledgments
|
|---|
B.R.H. is supported by a Population Health Award from the Alberta Heritage Foundation for Medical Research and a New Investigator Award from the Canadian Institute of Health Research. C.E.L. is supported by a New Investigator Award from the Canadian Institute of Health Research.
We are indebted to Kerri Gallo and Paul Heidenheim for data collection and statistical support.
 |
Footnotes
|
|---|
Published online ahead of print. Publication date available at www.cjasn.org.
Received March 30, 2006.
Accepted May 18, 2006.
 |
References
|
|---|
- 2001 Report, Volume 1: Dialysis and Transplantation, Ottawa, Ontario, Canadian Organ Replacement Register, Canadian Institute for Health Information,2001
- Reddan D, Klassen P, Frankenfield DL, Szczech L, Schwab S, Coladonato J, Rocco M, Lowrie EG, Owen WF Jr; National ESRD CPM Work Group: National profile of practice patterns for hemodialysis vascular access in the United States.
J Am Soc Nephrol13
:2117
2124,2002[Abstract/Free Full Text]
- Besarab A, Adams M, Amatucci S, Bowe D, Deane J, Ketchen K, Reynolds K, Tello A: Unraveling the realities of vascular access: The Network 11 experience.
Adv Ren Replace Ther7
:S65
S70,2000[Medline]
- Dhingra RK, Young EW, Hulbert-Shearon TE, Leavey SF, Port FK: Type of vascular access and mortality in US hemodialysis patients.
Kidney Int60
:1443
1451,2001[CrossRef][Medline]
- Mendelssohn DC, Ethier J, Elder SJ, Saran R, Port FK, Pisoni RL: Haemodialysis vascular access problems in Canada: Results from the Dialysis Outcomes and Practice Patterns Study (DOPPS II).
Nephrol Dial Transplant21
:721
728,2006[Abstract/Free Full Text]
- NKF-K/DOQI clinical practice guidelines for vascular access: Update 2000.
Am J Kidney Dis37
:S137
S181,2001[Medline]
- OMara N, Ali S, Bivens K, Shermann RA, Kapoian T: Efficacy of tissue plasminogen activator for thrombolysis in central venous dialysis catheters.
Hemodial Int7
:130
134,2003[Medline]
- Spry LM: Low-dose tPA for hemodialysis catheter clearance.
Dial Transplant30
:10
12,2001
- Eyrich H, Walton T, Macon EJ, Howe A: Alteplase versus urokinase in restoring blood flow in hemodialysis-catheter thrombosis.
Am J Health Syst Pharm59
:1437
1440,2002[Abstract/Free Full Text]
- Castner D: The efficacy of reteplase in the treatment of thrombosed hemodialysis venous catheters.
Nephrol Nurs J28
:403
410,2001[Medline]
- Nguyen TV, Dikun M: Establishing an alteplase dosing protocol for hemodialysis-catheter thrombosis.
Am J Health Syst Pharm61
:1922
1924,2004[Free Full Text]
- Falk A, Samson W, Uribarri J, Vassalotti JA: Efficacy of reteplase in poorly functioning hemodialysis catheters.
Clin Nephrol61
:47
53,2004[Medline]
- Macrae JM, Loh G, Djurdjev O, Shalansky S, Werb R, Levin A, Kiaii M: Short and long alteplase dwells in dysfunctional hemodialysis catheters.
Hemodial Int9
:189
195,2005[CrossRef][Medline]
- Hilleman DE, Dunlay RW, Packard KA: Reteplase for dysfunctional hemodialysis catheter clearance.
Pharmacotherapy23
:137
141,2003[CrossRef][Medline]
- Hamond J, Shalansky K, Jastrzebski J: Efficacy of low-dose alteplase for treatment of hemodialysis catheter occlusions.
J Vasc Access76
82,2005
- Manns B, Tonelli M, Yilmaz S, Lee H, Laupland K, Klarenbach S, Radkevich V, Murphy B: Establishment and maintenance of vascular access in incident hemodialysis patients: A prospective cost analysis.
J Am Soc Nephrol16
:201
209,2005[Abstract/Free Full Text]
- Daugirdas JT: Second generation logarithmic estimates of single-pool variable volume Kt/V: An analysis of error.
J Am Soc Nephrol4
:1205
1213,1993[Abstract]
- National Kidney Foundation: K/DOQI clinical practice guidelines for hemodialysis adequacy 2000.
Am J Kidney Dis37[Suppl]
:S7
S64,2001[Medline]
- Di Filippo S, Pozzoni P, Manzoni C, Andrulli S, Pontoriero G, Locatelli F: Relationship between urea clearance and ionic dialysance determined using a single-step conductivity profile.
Kidney Int68
:2389
2395,2005[CrossRef][Medline]
- Petitclerc T, Bene B, Jacobs C, Jaudon MC, Goux N: Non-invasive monitoring of effective dialysis dose delivered to the haemodialysis patient.
Nephrol Dial Transplant10
:212
216,1995[Abstract/Free Full Text]
- Lindsay RM, Bene B, Goux N, Heidenheim AP, Landgren C, Sternby J: Relationship between effective ionic dialysance and in vivo urea clearance during hemodialysis.
Am J Kidney Dis38
:565
574,2001[Medline]
- McIntyre CW, Lambie SH, Taal MW, Flucj RJ: Assessment of haemodialysis adequacy by ionic dialysance: Intra-patient variability of delivered treatment.
Nephrol Dial Transplant18
:559
563,2003[Abstract/Free Full Text]
- Mercadal L, Petitclerc T, Jaudon MC, Bene B, Goux N, Jacobs C: Is ionic dialysance a valid parameter for quantification of dialysis efficiency?
Artif Organs22
:1005
1009,1998[CrossRef][Medline]
- Petitclerc T, Goux N, Reynier AL, Bene B: A model for non-invasive estimation of in vivo dialyzer performances and patients conductivity during hemodialysis.
Int J Artif Organs16
:585
591,1993[Medline]
- Lindsay RM, Bradfield E, Rothera C, Kianfar C, Malek P, Blake PG: A comparison of methods for the measurement of hemodialysis access recirculation and access blood flow rate.
ASAIO J44
:62
67,1998[Medline]
- Hemmelgarn BR, Moist L, Pilkey RM, Lok C, Dorval M, Tam PY, Berall MJ, LeBlanc M, Toffelmire EB, Manns BJ, Scott-Douglas N; Canadian Hemodialysis Catheter Working Group: Prevention of catheter lumen occlusion with rT-PA versus heparin (Pre-CLOT): Study protocol of a randomized trial.
BMC Nephrol7
:8
,2006[CrossRef][Medline]
- Marshall MR, Golper TA, Shaver MJ, Alam MG, Chatoth DK: Urea kinetics during sustained low-efficiency dialysis in critically ill patients requiring renal replacement therapy.
Am J Kidney Dis39
:556
570,2002[CrossRef][Medline]
- Suri R, Depner TA, Blake PG, Heidenheim AP, Lindsay RM: Adequacy of quotidian hemodialysis.
Am J Kidney Dis42
:42
48,2003[CrossRef][Medline]
- Manzoni C, Di FS, Corti M, Locatelli F: Ionic dialysance as a method for the on-line monitoring of delivered dialysis without blood sampling.
Nephrol Dial Transplant11
:2023
2030,1996[Abstract/Free Full Text]
- Carson RC, Kiaii M, Macrae JM: Urea clearance in dysfunctional catheters is improved by reversing the line position despite increased access recirculation.
Am J Kidney Dis45
:883
890,2005[CrossRef][Medline]
- Depner TA: Assessing adequacy of hemodialysis: Urea modeling.
Kidney Int45
:1522
1535,1994[Medline]
- Clark WR, Leypoldt JK, Henderson LW, Mueller BA, Scott MK, Vonesh EF: Quantifying the effect of changes in the hemodialysis prescription on effective solute removal with a mathematical model.
J Am Soc Nephrol10
:601
609,1999[Abstract/Free Full Text]
- Jindal K, Chan CT, Deziel C, Hirsch D, Soroka SD, Tonelli M, Culleton BF: Chapter 4: Vascular access.
J Am Soc Nephrol17[Suppl 1]
:S16
S23,2006
- Twardowski ZJ, Haynie JD: Measurements of hemodialysis catheter blood flow in vivo.
Int J Artif Organs25
:276
280,2002[Medline]
- Sands J, Glidden D, Jacavage W, Jones B: Difference between delivered and prescribed blood flow in hemodialysis.
ASAIO J42
:M717
M719,1996[Medline]
- Stragier A, Wenderickx D, Jadoul M: Blood flow displayed by dialysis machines: Is it accurate?
EDTNA ERCA J22
:3
6,1996[Medline]