CJASN
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Published ahead of print on July 6, 2006
Clin J Am Soc Nephrol 1: 960-964, 2006
© 2006 American Society of Nephrology
doi: 10.2215/CJN.00810306

This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
CJN.00810306v1
1/5/960    most recent
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Lindsay, R. M.
Right arrow Articles by Sargent, K.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Lindsay, R. M.
Right arrow Articles by Sargent, K.

Dialysis

Hemodialysis Blood Access Flow Rates Can Be Estimated Accurately from On-Line Dialysate Urea Measurements and the Knowledge of Effective Dialyzer Urea Clearance

Robert M. Lindsay*, Jan Sternby{dagger}, Bo Olde{dagger}, Roland Persson{dagger}, Mary Ellen Thatcher*, and Kim Sargent*

* Optimal Dialysis Research Unit, London Health Sciences Centre, London, Ontario, Canada; and {dagger} Gambro AB Therapy Systems Research, Lyon, France, and Lund, Sweden

Address correspondence to: Dr. Robert M. Lindsay, London Health Sciences Centre, 800 Commissioners Road East, London, Ontario, Canada N6A 4G5. Phone: 519-685-8349; Fax: 519-685-8395; robert.lindsay{at}lhsc.on.ca

Measurement of blood flow rate (Qa) is used to monitor arteriovenous fistulas and grafts that are used for hemodialysis blood access. Most Qa measurements use indicator dilution techniques to measure the recirculation that is induced by the reversal of hemodialysis blood lines. R plus the dialysis circuit flow (Qb) allows the calculation of Qa. The principle of needle reversal also can be used with a dialysate urea monitor (e.g., DQM 200 [Gambro]) without injection of diluent; the effect of the reversal on urea concentration is observed. Access blood water flow rate (Qaw) in relation to the effective clearance (K) is found from the urea concentrations in the dialysate with needles in the normal (Cn) and reverse (Cr) positions: K/Qaw = (Cn – Cr)/Cr. Qa is calculated by adjusting Qaw for hematocrit and protein. For testing of this theoretical relationship, 20 patients who were dialyzed on Integra (Hospal) and Centrysystem 3 (Cobe) machines that were fitted with DQM 200 were studied. During each treatment, lines were reversed and Qa was measured by ultrasound velocity dilution (Transonic HD01 monitor); at the same time, Cn and Cr were measured by DQM 200 and K was calculated. K1 was determined from a predialysis blood urea concentration (Cb), initial dialysate urea concentration (Cd), dialysate flow rate (Qd), and the relationship K x Cb = Qd x Cd (K1). K was determined separately from a conductivity step method using Diascan (Hospal) attached to Integra machines only (K2). With the use of K1, 127 comparisons were made; a correlation existed (r = 0.916), although Bland-Altman analysis showed that the dialysate urea method gave a mean value 5.3% ± 15.3 (±SD) higher than that of Transonic (P < 0.001). With the use of K2, there also was a correlation of (r = 0.944; n = 63), and Bland-Altman testing showed an NS difference of +3.5% between the dialysate urea and Transonic methods. Qa can be estimated from on-line dialysate urea measurements that are taken before and after line reversal together with knowledge of K.